for Healthdyne Wallaby and Ohmeda Billiblanket

A.   Place the head of a dosimeter directly on the fiberoptic panel with the sensing surface facing the emitted light

B.  A minimum average of 6-8 mW/sqcm/nm must be obtained for the panel to be considered effective. More specific minimums follow below.

 

Minimum irradiance readings

Ohmeda Billiblanket at low setting:

6-8 mW/sqcm/nm

Ohmeda Billiblanket at high setting:

14-20 mW/sqcm/nm

Healthdyne model MD420, MD500, EG2000R:

6-8 mW/sqcm/nm

Healthdyne model MD420-DQL:

8-10 mW/sqcm/nm

Healthdyne model WPS-II, EG2000 at Level 1:

8 mW/sqcm/nm

Healthdyne model WPS-II, EG2000 at Level 2:

14 mW/sqcm/nm