IVAC Signature 7100
IVAC Signature 7100 and 7200
The IVAC Signature 7100 and 7200 infusion pumps are linear peristaltic devices that can be used for general IV infusions, antibiotics, blood infusions, enteral infusions, TPN and hyperalimentation, and lipids. The 7100 is a single-channel device, while the 7200 is a dual-channel device with independent programming on each channel. Monitoring options provide increased awareness of occlusion detection and battery charge condition. Dual-rate piggyback function and panel lock.
A multi-step program function allows a sequential program to deliver up to nine rates and volumes; can also be programmed using volume and time. Multiple infusions can be preprogrammed over a period of up to 24 hours; delay start and loading dose features. Computer interface for remote monitoring. Rotating pole clamp for pole or bed-rail mounting. Drug-specific dose rates can be calculated by selecting a drug name, which will then display on the screen. Recommended for epidural infusions. Optional flow sensor.
| Flow rate range: | 0.1-999.9
ml/hr in 0.1 ml increments (primary mode) 0.1-270.0 ml/hr in 0.1 ml increments (secondary mode) |
| VTBI range: | 0.1-9999.9
ml in 0.1 ml increments (primary mode) 0.1-999.9 ml in 0.1 ml increments (secondary mode) |
| KVO range: | 0.1-20.0 ml/hr in 0.1 ml increments |
| Battery Life: | New, fully-charged battery lasts approx. 4 hours at 100 ml/hr. for a 2-channel instrument operating on both channel simultaneously. |
| Recharge Time: | Recharges to 95% capacity in 4 hours while operating. |
| PM Interval: | Manufacturer recommends 1 year PM interval. |
| Occlusion Pressure Limit: | 12 ± 4 PSI in pressure mode. |
| Accuracy: | ± 5 % |
Administration sets: Use only IVAC 72 series administration sets.