Sabratek Infusion Devices

Sabratek infusion pumps include the pole-mounted Sabratek 3030 and the ambulatory, multi-therapy Sabratek 6060 Homerun. Both devices are capable of performing continuous, auto-ramp (automatic, linear tapering), delayed start, intermittent, and sequential delivery of varying rates and volumes. In addition, the 6060 Homerun is capable of PCA infusion, either continuous, continuous with bolus, or demand bolus only. Both can be made tamper-resistant by requiring the use of a password to enter the programming mode.


Sabratek 3030 and the Sabratek 6060 Homerun

Sabratek 3030

The Sabratek 3030 is a pole-mounted linear peristaltic device that can be used for either venous or arterial infusions. Unlike the Sigma 6000+P, which is factory-calibrated to use standard straight-line tubing from either McGaw, Abbott, or Baxter, the 3030 can be calibrated by the user for any of the three manufacturer’s compatible sets, along with sets manufactured by Sabratek (see the attached list of compatible sets). The pump can deliver medication in four different ways:

1) Continuous - a continuous delivery at a single programmed flow rate;

2) Auto-ramp - automatic, linear ramping (tapering);

3) Intermittent - delivery of a preset number of doses at regular intervals; and

4) Sequential - sequential delivery of up to 15 different rates and volumes.

The pump can be programmed using either rate information (milliliters/hour) or using delivery time (volume to be infused over a programmed delivery time). The pump can also delivery Secondary (Piggyback) infusions before or during the primary infusion, and has the ability to titrate (infusion rate can be changed without halting the infusion). The infusion can be delayed to start at a later time if required. The 3030 has an 2-position adjustable pressure setting using either a low or high pressure threshold.

If the pump is turned off before the infusion is complete, the memory will retain rate, VTBI, and volume infused information; when the pump is restarted, it will continue from the point at which it was interrupted. Programmed infusion settings are maintained in the memory indefinitely, until they are changed by the operator. The pump requires a Sabratek 3030 flow sensor to operate with non-Sabratek sets.

Flow Rate: 1 - 999 ml/hr. in 1 ml. increments
Volume Limit: 1 - 9999 ml. in 1 ml. increments
Accuracy: ± 5 %
Weight: approximately 10.5 lb. (4.5 kg)
Battery: 12-volt, rechargeable, will run approx. 6-7 hours at 125 ml/hr.
"LO BATTERY" indicates 60 minutes of power remaining
Battery recharge: approx.24 hrs for discharged battery while infusing on AC power.
Occlusion: LO setting: 8 ± 5 psi
HI setting: 20
± 5 psi
KVO rate: 1 ml/hr.
1 - 9 ml/hr. in Intermittent mode.
Memory: Retains programming settings indefinitely until changed by operator.

Sabratek 6060 Homerun

 

The Sabratek 6060 Homerun is an ambulatory multi-therapy device that can be used for enteral, epidural, subcutaneous, arterial and intravenous applications, using rotary peristaltic operation. It has five different methods of delivery:

1) Continuous - continuous delivery of a single rate and volume;

2) Auto-ramp - automatic, linear ramping (tapering);

3) Intermittent - delivery of a preset number of doses at regular programmed intervals;

4) Sequential - sequential delivery of up to 25 different rates and volumes; and

5) PCA - continuous only, continuous with demand bolus dose, or bolus dose only.

The 6060 Homerun can be programmed using either milligrams or milliliters, and provides the option of programming using rate (ml/hr.) or time (volume delivered over a specific time period). Memory is indefinite until parameters are changed by the user, and if the pump is stopped during delivery, it will begin at the same point in the infusion when restarted. The pump uses YES/NO response to on-screen questions to facilitate easy programming.

The pump displays a flashing arrow on screen to indicate when it is infusing; in low ambient light situations, pressing a key during operation will activate a screen back-light for ten seconds to view the display.

The Homerun also has six standard data settings that are user-adjustable: audio volume, KVO rate, occlusion pressure (low or high), display language (English or Spanish), air detection (size of air bubble), and upstream occlusion (enabled or disabled).

The outer care is impact-resistant and has water-resistant enclosure seals to reduce fluid penetration. The pump also uses a remote bolus cord which can be used instead of the keypad bolus button. The pump uses only Sabratek 6060 administration sets (see accompanying list). The pump includes a clear hard plastic "SabraSafe" fluid bag holder which attaches to the head of the pump; it holds the fluid bag inside, while the fluid can be inspected through the clear case for particles.

Flow rate: 0.1-400 ml/hr. using standard cassette; 0.1 - 50.0 ml/hr. w/ microcassette
Volume limit: 0.1-9999 ml. (standard volume); 0.1-999 ml. (PCA volume)
Accuracy: ± 6 %
Weight: 13 oz. (350 g.)
Power Source: Two 9-volt alkaline or lithium batteries, external batt pack, or AC Batt. Eliminator

Sabratek 3030 Administration Sets (all 30/cs)

Basic No Y-site Sets
with On/Off Clamp:
SR500000
SR500012
SR500022
Basic Sabraset, 105"
Basic Sabraset w/1.2 micron filter, 105"
Basic Sabraset w/.22 micron filter, 105"
Basic One Y-site Sets
with On/Off Clamp:
SR500100
SR500112
SR500122
Basic Sabraset, 105"
Basic Sabraset w/1.2 micron filter, 105"
Basic Sabraset w/.22 micron filter, 105"
Primary Two Y-site Sets with
Drip Chamber and Roller Clamp:
SR500200
SR500212
SR500222
SR500300
Primary Sabraset, 105"
Primary Sabraset w/1.2 micron filter, 105"
Primary Sabraset w/.22 micron filter, 105"
Secondary Sabraset with IV Hook, 25"

Sabratek 6060 Administration Sets (all 30/cs)

SR560100 78" IV Set with Cassette, Spike, Slide Clamp, & Male Luer
SR560112 84" IV Set with Cassette, Spike, Slide Clamp, 1.2 m filter and Male Luer
SR560200-100 13" IV Set with Cassette, 100 ml container, Slide Clamp & Male Luer
SR560300 53" Extension Set with Slide Clamp and Male/Female Luer
SR560322 59" Extension Set with .22 m filter, Slide Clamp, & Male/Female Luer
SR560500-100 63" IV Set with Cassette, 100 ml container, Slide Clamp, & Male Luer
SR560522-100 69" IV Set with Cassette, 100 ml container, .22 m filter, Slide Clamp, & Male Luer
SR567100 Non-DEHP set for Taxol/Nitroglycerin infusion

6060 Accessories and Replacement Parts

606010 SabraSafe Fluid Bag Holder  
606600 Fanny Pack w/o Logo
606030 Hard Carrying Case  
606600-1 Fanny Pack w/Logo
606050 AC Battery Eliminator  
606601 Slim Pack w/o Logo
606060 External Battery Pack  
60601-1 Slim Pack w/Logo
606070 SabraNest Mini Pole Clamp 
606121 Battery Door  
606602  Back Pack w/o Logo
606602-1 Back Pack w/ Logo
606380 Remote PCA Cord