Medfusion 2001 & 2010 Syringe Infusion Pumps

Medfusion 2001

The Medfusion 2001 & 2010 are ambulatory syringe pumps that run on either AC current or on rechargeable internal batteries (on-screen voltage test is available). Each accepts a broad range of syringe brands and sizes, and has automatic sensing of syringe size. The 2010 is limited to either continuous infusion or Volume/Time programming, but it has the capability of both bolus infusion and of programming by body weight (mcg/kg/min, mcg/kg/hr, mg/kg/min, or mg/kg/hr), by mass (mg/hr), or by volume (ml/hr). The 2001 has no bolus function, but it can be programmed for continuous, auto or manual intermittent, or Volume/Time infusions. Both are recommended for intravenous or arterial infusions of blood, lipids, fluids, antibiotic, and any medication requiring precisely-controlled infusion rates, including epidural infusions. The pumps have 4-position variable occlusion settings.

A custom program mode allows preselection of certain modes and parameters, which can then be locked-out by the programming clinician to limit program access. The syringe manufacturer can also be custom-programmed if not using a syringe by BD, Monoject, Abbott (50cc only), or Terumo. Both models are also indicated for use with BD glass syringes (1, 3, 5, or 10cc sizes). In addition, software version 1.4A of the 2001 can also use all Braun syringe, 10 and 50cc syringes by Fresenius, and 50cc Perfusor syringes.

Both pumps can also be pole-mounted, using the pole clamp included with the pump. Each has an easy-to-read LCD screen with LED alarm and function lights. Accuracy is ± 3 %, excluding variations in syringe accuracies.

Flow rate: 0.1 - 378.0 ml/hr., depending on syringe size selected
Volume limit: 00.1 to max. capacity of syringe in increments of 0.1 ml
Vol. delivered: (2001) - 000.00 to 999.99 in increments of 0.01 ml
(2010) - 0000.0000 to 9999.9999 mg or 0000.00 to 9999.99 ml
Bolus rate: (2010 only) - Up to 378.0 ml/hr
KVO: 0.1 - 9.9 ml/hr in increments of 0.1 ml
Occlusion: Normal: max. 16.4 PSI with Mono 35cc syringe
Low: max. 5.96 PSI with Mono 35cc syringe
Middle: max. 13.8 PSI with Mono 35cc syringe
High: max. 15.0 PSI with Mono 35cc syringe
Accuracy: ± 3 %, excluding variations in syringe accuracies.
Battery capacity: 16-hour charge operates the the pump for approx. 10 hours at 5.0 ml/hr with a 60-ml syringe
Recharge time: 16 hours maximum

Return to Syringe Index