Medfusion Ambulatory Infusion Pumps
Medfusion manufactures a line of lightweight ambulatory infusion devices known as the WalkMed series, along with a larger but still ambulatory series of pumps, the Medfusion 2001 and 2010. WalkMeds are small, square devices which utilize linear peristaltic operation and run on a single 9-volt battery. They carry a small 65ml reservoir in the back cover of the pump for infusion of minimal volumes. Larger volumes can be infused using the 150 or 250 ml reservoirs, carried in the optional 250 ml pouch. The WalkMeds require either the standard set MFPS-400/300 with any M/F extension set, or the PS-360, which includes a 60" extension set. The SP-6 Spike Set can also be used for larger infusions in conjunction with a medication container with spike input.
The popular and easy-to-program Medfusion InfuMed 300 pump has been discontinued (parts and service support are no longer offered) and replaced with the WalkMed 350. Both offer continuous infusion, with the 300 programming in tenths of a ml, the 350 in tenths or hundredths of a ml.
The 400 series of WalkMeds are specialized pumps, each one having a therapy focus: the 410 is continuous only, the 420 I/C offers continuous and intermittent infusion, the 430 PCA and WalkMed PCA have continuous and PCA infusion, and the 440 PIC (replacing the discontinued WalkMed 400) performs continuous, intermittent, and PCA.
All the WalkMeds have an accuracy of ± 5 %.The Medfusion 2001 and 2010 are syringe pumps which can utilize almost any commercially-available syringe and syringe size. They can be used as stationary pole-mounted devices or as ambulatory units. Both run on rechargeable internal batteries and are recharged by an external AC adapter. The 2001 and 2010 both have an accuracy of ± 3 %, not counting syringe variance.
|
|
Clockwise, from top left: |
Medfusion Infumed 300
MF300 Features: Linear peristaltic operation, simple to program, On/Off/Prime 3-position switch, palm-size, operates on 9-volt alkaline battery. 65mL reservoir fits inside back cover, can use any collapsible IV bag with MF SP-6 Spike set adaptor. Rotary switches for rate program entry. Optional patient Lock-out kit is available (ILK-101).
Applications: Chemotherapy, Pain Management, and other continuously-infused medications.
Rate Parameters: 0.1 - 9.9 mL/hr in increments of 0.1 mL/hr
Battery Life: 165 hours at 5.0 mL/hr rate; 520 hours at 0.5 mL/hr rate
Accuracy: ± 5 %
Alarms: Audio and LED for Low Battery, Depleted Battery, Occlusion, System Malfunction, Prime Mode. Normal operation gives blinking green LED light.
Occlusion: 10 ± 3 PSI
Medfusion Infumed 350
Applications: Chemotherapy, antibiotics, analgesics, and any other drug requiring continuous intravenous, intra-arterial, or subcutaneous infusion rates. Epidural infusions delivered up to 96 hours.
Features: Linear peristaltic operation, simple to program, digital input of program rates with LCD display, optional patient lock-out kit, 65-mL reservoir fits inside back cover, any size collapsible IV bag can be used with MF SP-6 Spike set adaptor. Uses 9-volt alkaline. Replaces the MF 300, no longer available.
Rate Parameters: 0.10 - 19.99 mL/hr in increments of 0.01
Battery Life: 450 mL at 1 mL/hr; 650 mL at 10 mL/hr
Accuracy: ± 5 %
Medfusion Walkmed 400, 400PCA, 410C, 420I/C, 430PCA, 440PIC
400, 440 PIC: Continuous, Intermittent and PCA
400, 430 PCA: Continuous, and PCA Infusion
410 C: Continuous Infusion
420 I/C: Continuous and Intermittent Infusion
Features: Linear Peristaltic operation, Digital program input with LCD Display,. Programmable concentration. Lock levels to prevent patient tampering; Total Volume Limit; KVO Rate can be programmed if Intermittent Mode available; PCA 24-hour history recall and remote bolus button. Runs on single 9-volt alkaline battery. Can be programmed in tenths or hundredths. Programmable Near-End Alarm.
Continuous
Flow Rate: 0.1 - 30.0 mL/hr in increments of 0.1 or 0.01 - 19.99 mL/hr in increments of 0.01
Total Volume Delivered: 0 - 1999 mL
Total Volume Limit: 0 - 1999 mL
Intermittent
Delivery Time: 1 minute - 19 hours, 59 minutes
Delivery Volume: 0.1 mL - 199.9 mL
Time Between Delivery: 1 min. - 19 hours, 59 min
Total Volume Delivered: 0 - 1999 mL
Volume Limit: 0 - 1999 mL
KVO Rate: 0 - 9.9 mL/hr
PCA
Concentration: 0.1 - 9.9 mg/mL or 1 - 99 mg/mL
Basal Rate: 0 - 199.9 mg/hr or 0 - 1999 mg/hr or delivery in mL
Bolus Dose: 0 - 199.9 mg or in mL
Lockout Time: 0 - 99 minutes
One-Hour Volume Limit:1 - 1999 mg/hr or in mL
Total Volume Limit: 0 - 1999 mg or in mL
Alarms & Alerts: Occlusion, Volume Limit (in mg & mL), One-hour volume limit (PCA only), Under-delivery, Over-delivery, System Malfunction, Bar Open Alarm, Delivering, Pump Left in Program Mode, Program Error, Prime, Total Volume Delivered (in mg & mL), Near-End alarm (programmable), Low Battery, Depleted Battery.
Accuracy: ± 5 %
Occlusion: 10 ± 3 PSI