Excelsior ESP Syringe Pumps

The Excelsior ESP Syringe Pump is similar in appearance and operation to the Baxter 300XL. Like the 300XL, it features three motor speeds which allow the clinician to set the time of delivery, and accepts most syringe sizes. Using the plug-in adapter (shown at left), the pump can also infuse solutions from 140 ml syringes which are available from Excelsior or Monoject. The adapter contains electronic components inside the plastic piece which suplement the internal pump motor electronics and allow the larger infusion; an extra-large syringe clamp will accept syringes of 3ml. up to the 140 ml. size. The time of infusion is also longer than the 300XL’s; in the Xslow mode, infusion time can be as long as 150 minutes.
Rate Parameters: 5.2 - 77.5 ml/hr.
Time Parameters: 10-150 minutes
Syringe Sizes: Monoject, Terumo, BD, or Excelsior, from 3ml - 140 ml.
Occlusion Limits: 9.2 - 45 PSI, depending on syringe
Batteries: four size "C" cells
Battery Life: 400 hours typical
Accuracy: ± 3 %

Return to Syringe Index