Sims Deltec (Pharmacia) Ambulatory Infusion Devices
Administration Sets and Accessories
| Left to Right: CADD-1 (5100), CADD PCA II (5800), CADD+ (5400) |
Sims Deltec CADD-1 (5100)
The Sims Deltec CADD 1 pump delivers in mls per 24 hours instead of the usual ml/hour of the other CADD models, but can also deliver 90 ml/hour in the fixed high-flow mode. It can be used for any continuous therapy, and is often recommended for chemotherapy treatment.
Delivery Modes: Continuous
Rate Parameters: 0-299 mL/24 hrs, or 90 mL/hr in fixed high-flow mode
Accuracy: ± 6 %
Hi-Pressure Alarm: 16-40 PSI
CADD+ (5400)
The Sims Deltec CADD+ pump delivers in two modes, either continuous or intermittent. Because of its intermittent capability, it is often recommended for the delivery of antibiotics, which requires a delivery period followed by a KVO (Keep Vein Open) rest period, and then repeats the same program a number of times. The CADD + can also be use in the continuous mode, delivering up to 75 mls/hour. There is no screen for the entry of drug concentration on the CADD+.
Recommended Uses: Intermittent Delivery of Antibiotics, Continuous Delivery of Chemotherapy
Delivery Modes: Continuous or Intermittent
Rate Parameters: 0-75 mL/hr
Reservoir Volume: Off, 1-1500 mL
Infusion Volume: 0-150 mL
Infusion Period: 10 minutes - 12 hours
Infusion Cycle: 20 minutes - 12 hour 10 minutes or 24 hours
KVO Rate: 0-10 mL/hr
Delay Start: 10 minutes - 12 hours
Accuracy: ± 6 %
Hi-Pressure Alarm: 16-40 PSI
CADD PCA II (5800) and CADD PCA Remote (5800R)
| Recommended Uses:
Pain Management, any continuous infusion Delivery Modes: Continuous Delivery, Patient or Clinician-activated Bolus; Programmable concentration (0-95 mg/mL) Rate Parameters: 0-20 mL/hr, or 0-99.5 mg/hr Patient Bolus: 0-6 mL or 0-99 mg Clinician Bolus: 0-20 mL or 99.5 mg Accuracy: ± 6 % Hi-Pressure Alarm: 16 - 40 PSI |
Sims Deltec CADD PCA and CADD PCA Remote Differs from the earlier (discontinued) CADD PCA (5200) in Clinician-activated bolus capability, Attempted Doses informational screen, and automatic delivery resumption after HI-P alarm when occlusion pressure is relieved (5200 required a restart after pressure alarm). The 5800R includes a remote bolus button which attaches to the side of the housing. Both are used for PCA infusion (Patient-Controlled Analgesia), which allows the patient to self-bolus as needed in addition to providing a continuous basal infusion of pain medication. |
Sims Deltec CADD TPN (5700)
Recommended Uses: TPN (Total Parenteral Nutrition)
Delivery Modes: Continuous Delivery or Continuous Delivery with Taper Up and/or Down
Rate Parameters: Using Power Pack: 10-400 ml/hr; Using 9-volt Alkaline Battery: 10-250 mL/hr
Reservoir Volume: Off, 10-9990 mL
Infusion Period: 10 minutes - 99 hours, 50 min.
Taper-up Period: 0 - 99 hours, 40 minutes
Taper-down Period: 0 - 99 hours, 40 minutes
KVO Rate: 5 mL/hr
Accuracy: ± 8 %
Hi-Pressure Alarm: 10 - 34 PSI
Power Pack Recharge: Up to 12 hours
Power Source Specifications:
| 9V Alk | 9V Lithium | Power Pack | |
| Maximum Rate | 250 mL/hr | 250 mL/hr | 400 mL/hr |
| Battery Life | 6 hours | 18 hours | 11 hours |
| Maximum Delivery | 1500 mLs | 4500 mLs | 4500 mLs |