AVI 280 RT (Rate/Taper) This pump operates in either a standard rate/volume mode similar to the 200A, or in the Rate/Taper mode, in which the keypad accepts maximum infusion rate and total volume infused information for TPN infusions; the pump calculates the proper taper-up, taper-down, and maintenance level delivery time periods. It requires the newer "8C"-prefix administration sets with automatic free-flow protection clips. Manufacture of this pump has been discontinued, but service is still supported. |
| Rate Parameters: | Standard mode: 1 - 999 ml/hr in 1 ml increments Rate Taper mode: 10 - 400 ml/hr in 1 ml increments |
| Volume Parameters: | 1 - 9999 ml in 1 ml increments |
| Battery capacity: | Approx. 8 hours at 125 ml/hr. |
| Battery recharge: | 14 hours, no operation; 24 hours operating at 125 ml/hr. |
| Memory: | Indefinitely stores all settings in memory when pump is shut off. If pump is shut off before a rate/taper program is finished, it will continue the infusion at the exact point where it left off when next turned on, unless cleared and restarted. |
| Accuracy: | +/- 2% |
| KVO: | 1 ml/hr. |
Compatible Administration Sets